Recall Z-0319-2024 ยท JAK System, X-Ray, Tomography, Computed

Philips North America

Spectral CT on Rails, Software Version 5.1.0, Model Number 728334

DateNov 16, 2023 (initiated Oct 9, 2023)
StatusOpen, Classified
ReasonThis recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This issue could result in the gantry moving to a location the operator does not expect. If this occurs, it presents a risk that the scan could occur at an incorrect location, or that the unintended motion of the gantry could cause a collision with the user or needle. The other part of the recall involves a software crash when the CCT foot pedal is activated (as described in RES ID9336
Root causeUnder Investigation by firm
ActionAn URGENT Medical Device Correction notification letter dated 10/9/23 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users Below are short-term precautions to take until the permanent solution is installed. Remain vigilant and confirm gantry position during procedures. Review the Emergency Stop Procedure and Collision Prevention Guidelines provided in Section 6: Working with Specialized Exams of the IFU. Actions planned by Philips to correct the problem Philips will contact you to schedule a time for a Philips Field Serv
Quantity1 unit
DistributionUS Nationwide distribution in the state of MN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K212875Spectral CT on Rails
Philips Medical Systems Nederland B.V., May 13, 2022