Recall Z-0495-2024 ยท HBG Drills, Burrs, Trephines & Accessories (Manual)
Integra LifeSciences Corp.
Integra Cranial Access Kit-For access to the subarachnoid space or the lateral ventricles of the brain. Catalog Number: INS7260
| Date | Dec 6, 2023 (initiated Oct 19, 2023) |
| Status | Open, Classified |
| Reason | Defect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device |
| Root cause | Process control |
| Action | Integra notified consignees (Hospitals, Government facilities, and Distributors) via Fedex priority overnight on 10/19/23 . Letter titled: 1st Notification- Urgent: Voluntary Medical Device Recall. Letter states reason for recall, health risk and action to take: Actions to be Taken by Customers: 1. If you do have units of the affected product listed in Table 1, remove it immediately from service. 2. If you do have affected product, check the box on the enclosed form I do have affected product. Record the total quantity of the affected product that you have. 3. If you do not have affected |
| Quantity | 510 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, Chile, France, Ireland, Israel, South Africa, United Kingdom. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K961113 | CNS INTERCHANGEABLE BIT CRANIAL DRILL Clinical Neuro Systems, LLC, Jun 5, 1996 |