ClearanceBrief
Recall Z-0493-2024 ยท HBG Drills, Burrs, Trephines & Accessories (Manual)

Integra LifeSciences Corp.

Integra Cranial Access Kit - For access to the subarachnoid space or the lateral ventricles of the brain. Catalog Number: INS7040

DateDec 6, 2023 (initiated Oct 19, 2023)
StatusOpen, Classified
ReasonDefect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device
Root causeProcess control
ActionIntegra notified consignees (Hospitals, Government facilities, and Distributors) via Fedex priority overnight on 10/19/23 . Letter titled: 1st Notification- Urgent: Voluntary Medical Device Recall. Letter states reason for recall, health risk and action to take: Actions to be Taken by Customers: 1. If you do have units of the affected product listed in Table 1, remove it immediately from service. 2. If you do have affected product, check the box on the enclosed form I do have affected product. Record the total quantity of the affected product that you have. 3. If you do not have affected
Quantity59 units
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, Chile, France, Ireland, Israel, South Africa, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K961113CNS INTERCHANGEABLE BIT CRANIAL DRILL
Clinical Neuro Systems, LLC, Jun 5, 1996