ClearanceBrief
Recall Z-0438-2024 ยท DWC Controller, Temperature, Cardiopulmonary Bypass

CardioQuip, LLC

CardioQuip Modular Cooler-Heater, Model:MCH-1000(m), Power: 115V-60hz,16A

DateDec 1, 2023 (initiated Oct 19, 2023)
StatusOpen, Classified
ReasonThere is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components.
Root causeDevice Design
ActionOn 10/19/23, CardioQuip issued a "Urgent: Medical Device Correction" notification to all affected consignees. This notification was an update to the firm's May 22, 2023 communication. CardioQuip ask consignees to take the following actions: 1. Prime the MCH-1000(i) and MCH-1000(m) prior to each use. 2.Never operate the device dry (see communication for further details). 3. Design Changes to be implemented (see communication for further details). A. Relocation of the water Level Sensor (see communication for further details). B. Software Update (see communication for further details). 4. Ple
Quantity150 units
DistributionUS Distribution: TX, IN, IL, WI, FL, NJ, PA, KY, VA, WV, CT, CA, WA, AL, IA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102147CARDIOQUIP MODULAR COOLER-HEATER MODEL MCH-1000
Cardioquip Llp, Nov 19, 2010