Recall Z-0405-2024 ยท NOU Continuous, Ventilator, Home Use

Breas Medical, Inc.

with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to ventilator device, Vivo 45LS. intended for partial insertion into a patient's mouth (in between the lips) to facilitate access to the respiratory system.

DateNov 27, 2023 (initiated Oct 13, 2023)
StatusOpen, Classified
ReasonProduced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing patients using MPV-VCV and MPV-PCV modes of ventilation
Root causeProcess control
ActionBreas Medical contacted customers via telephone followed by email on 10/13/23. Letter (Breas reference: SCAR-330) states reason for recall, health risk and action to take: Breas Medical recommends that all those potentially affected by this notice inspect existing inventory and identify mouthpieces which have been set up with the Vivo 45 LS for patient use. All defective mouthpieces shall be discarded. Breas Medical will provide replacement NIV Angled Mouthpieces, article number 008234, for all defective inventory and for devices in patient possession, where applicable. To obtain replacement N
Quantity215 units
DistributionNationwide Foreign: Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193586Vivo 45 LS
Breas Medical AB, Jan 27, 2021
K233452Vivo 45 LS
Breas Medical AB, Jan 13, 2025