ClearanceBrief
Recall Z-0449-2024 ยท EOQ Bronchoscope (Flexible Or Rigid)

AURIS HEALTH INC

Monarch Platform REF MON-000005-01R, 100-240V- 50/60Hz 1440w

DateDec 1, 2023 (initiated Oct 18, 2023)
StatusOpen, Classified
ReasonTheir is the potential that software issues may result in flipped image of the virtual Bronchoscope view.
Root causeSoftware design
ActionOn October 19, 2023, Auris Health, Inc. issued an "URGENT: MEDICAL DEVICE CORRECTION" notification to affected consignees. Auris asked consignees to take the following actions: Should an image flip be detected in the Virtual Bronchoscope view while navigating toward the target, please perform the following steps to correct the image orientation: 1. Retract the MONARCH Bronchoscope outside the patient 2. Undock the MONARCH Bronchoscope from the arms 3. Clear any fault pop-up messages 4. Re-dock the MONARCH Bronchoscope to the arms 5. Re-initialize Navigation and proceed with the case 6. Please
Quantity110 units In total
DistributionUS Distribution: AK, AL, AZ, CA, CO, CT, DC, DE, FL, GA, IL, IN, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OR, PA, SC, SD, TN, TX, WI & WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193534Monarch Platform
Auris Health, Inc., Apr 15, 2020