ClearanceBrief
Recall Z-0426-2024 ยท FNL Bed, Ac-Powered Adjustable Hospital

Stryker Medical Division of Stryker Corporation

ProCuity bed series, model number 3009, item number: 3009PX-L-200

DateDec 1, 2023 (initiated Aug 30, 2023)
StatusOpen, Classified
ReasonPatient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.
Root causeProcess control
ActionA customer notification was sent October 20, 2023 via FedEx. Users are to (1) Locate affected units and identify the address where they can be serviced. (2) Return the business reply form by fax +1 269 488-8691 or email to productfieldaction@stryker.com. Upon receipt of the completed business reply form, Stryker will contact you to arrange a time to complete the required electrical tests. (3) Forward notification to any new user/location and advise Stryker of the new address or device status if disposed of or obsolete using the reply form. Address questions or concerns to Customer Service at
DistributionDistribution US Nationwide including Puerto Rico, Australia, Canada, Chile, China, Mexico, Poland, Spain and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.