ClearanceBrief
Recall Z-0398-2024 ยท JAI Couch, Radiation Therapy, Powered

LEONI CIA CABLE SYSTEMS

ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in Radiation Therapy, Radiology, or Radiology image-guided treatment environment. Part Number: 1SOPT00003

DateNov 21, 2023 (initiated Oct 6, 2023)
StatusOpen, Classified
ReasonUsers of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the movements are again allowed after an unforeseen interruption signal of motion may result in a collision of the patient with another stationary component of the treatment room.
Root causeSoftware design
ActionBizLink s notified customer via email with delivery and read confirmation and with the Field Safety Notice (FSCA Ref: UMCU-638) on October 6, 2023. Letter states reason for recall, health risk and action to take: BizLink technical support team will contact customers shortly to organize the deployment of the hot fix in the proton-therapy treatment centers. While waiting for the hot fix, BizLink recommend that customers implement the following action at the customer software level to avoid this problem: Send a stopFast command just after reactivating the motionEnable / the E-stop / the collisio
Quantity1 unit
DistributionUS Nationwide distribution in the states of MA, TN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160518LEONI ORION System
Leoni Cia Cable System, Jul 14, 2016