Recall Z-0406-2024 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

Da Vinci Single-Site Wristed Needle Driver, REF: 478115, 5 mm

DateNov 28, 2023 (initiated Oct 6, 2023)
StatusOpen, Classified
ReasonThere is the potential for the needle drive to become dislodged and/or broken which could result in uncontrollable movement.
Root causeProcess control
ActionOn October 24, 2023, Intuitive issued a "Urgent Medical Device Recall notification" to all affected consignees via E.Mail. Intuitive asked consignees to take the following actions: 1. Read and understand the contents of the letter. 2. Locate and return all affected lots of 478115-03 in your inventory via the standard RMA process by calling Intuitive Customer Service based on your region below. 3. Notify all surgeons, personnel, and sites using da Vinci X and da Vinci Xi Single Site Wristed Needle Driver, PN 478115-03 that they should review and understand the contents of this letter. 4. Compl
Quantity317 units
DistributionUS: IL, FL, ID, SC, MI, TX, TN, NY, OR, CO, CA, MS, NJ, Puerto Rico OUS: South Korea
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152448da Vinci Single-Site Instrument and Accessories
Intuitive Surgical, Inc., Mar 9, 2016