Recall Z-0670-2024 ยท IQI Orthosis, Limb Brace

Townsend Design

CUSTOM SPRYSTEP, custom made device dynamic ankle-foot orthosis (AFO) , Ref #s: 17C1000, 17C2000, 17C1010, 17C2010, 17C1040, 17C2040

DateJan 5, 2024 (initiated Sep 23, 2023)
StatusCompleted
ReasonDue to potential air pockets between layers of material, their is the potential for complete struct breakage of knee-ankle-foot orthosis.
Root causeProcess control
ActionOn November 7, 2023, Townsend Design (Thuasne USA), issued a "Urgent Medical Device Recall notification to affected consignees: Townsend Design asked consignees to take the following actions: 1. Clinician should contact any patient who was fit with a device received between June 14th and September 10, 2023 to notify them of the recall and schedule a visit to examine the patient's device. 2. During the patient visit, the clinician should examine the device for structural defects (delamination, cracking, complete breakage, etc.) in the area identified in the picture above in Section 3 3. If
Quantity248 units
DistributionUS: AL AR AZ CA CO CT DC DE FL GA IA ID IL IN KENT KS KY LA MA MD ME MI MN MO MS MT NC NE NH NJ NM NV NY OH OK OR PA QC SC SD TN TX UT VA VT WA WI WV WY, Puerto Rico, OUS: None OUS: Canada, New Zealand, Japan, Australia, Netherlands, United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.