ClearanceBrief
Recall Z-0323-2024 ยท MQV Filler, Bone Void, Calcium Compound

Musculoskeletal Transplant Foundation, Inc.

DBX Demineralized Bone Matrix Putty 10cc-Indicated for treatment of surgically created osseous defects or osseous defects created from traumatic injury. Model Number: 038100

DateNov 17, 2023 (initiated Sep 22, 2023)
StatusOpen, Classified
ReasonUnits are labeled as 10cc; however the units have a fill volume of 1cc units. This may result in prolongation of surgery.
Root causeLabeling Change Control
ActionMTF Biologicals issued Voluntary Market Withdrawal Notification Letter via customer service scripts on 9/22/23 and provided written notification via letter and/or Email on 10/3/23 to customers . Letter states reason for recall, health risk and action to take: If the tissue is still available in your inventory, we request you return the tissue to us. We will send a replacement if required. Should you have any questions, please contact Customer Service at 1-800-433-6576, or please feel free to contact me at 732-661-2209, or email MTF Regulatory Affairs at Regulatory_Customer_Inquiry@mtf.org.
Quantity18 units
DistributionUS Nationwide distribution in the states of PA, TX, WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040262DBX DEMINERALIZED BONE MATRIX PUTTY, BONE MATRIX PASTE AND BONE MATRIX MIX
Musculoskeletal Transplant Foundation, Mar 17, 2005