ClearanceBrief
Recall Z-0402-2024 ยท EWO Audiometer

PATH

SENTIERO Screening2, mobile medical devices for hearing examinations, Model Number SOH100098, Part Number 100098-S2

DateNov 25, 2023 (initiated Sep 26, 2023)
StatusOpen, Classified
ReasonAffected lots of product may experience technical distortions to a greater extent than expected.
Root causeProcess change control
ActionAn URGENT: MEDICAL DEVICE RECALL SENTIERO notification letter dated 9/26/23 was sent to customers. Actions to be taken by the User: The technical distortions can influence the specified performance of the device when performing diagnostic DPOAEs at high stimuli. Therefore, PATH MEDICAL chose to voluntarily recall affected products for correction. 1. Immediately identify any affected devices at your location and quarantine them. 2. Please return your affected device(s) to your local distributor or PATH MEDICAL directly (Shipping Address: Landsberger Str. 65, 82110 Germering, Germany) to have y
Quantity28 units
DistributionUS States: FL, MO, OH, WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133012SENTIERO
Path Medical GmbH, Jun 9, 2014
K213345Sentiero, AuDX, AuDX PRO, Sentiero Desktop, AuDX PRO Flex, Sentiero Advanced, NavPRO ONE
Path Medical GmbH, Jun 22, 2022