Recall Z-0352-2024 ยท GHH Fibrin Split Products

Roche Diagnostics Operations, Inc.

Tina-quant D-Dimer Test System, Fibrinogen/Fibrin Degradation Products Assay, Material Numbers 03001245322 - (cobas c 311, 501 and 502; INTEGRA 400 plus), 08105618190 - (cobas c 303 and 503)

DateNov 20, 2023 (initiated Oct 3, 2023)
StatusOpen, Classified
ReasonElevated results were detected with Li-heparin plasma samples when compared to citrated plasma samples.
Root causeUnder Investigation by firm
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 10/3/23 was sent to customers. Actions Required " Roche recommends using citrated plasma for D-Dimer quantification until further notice. " If your laboratory continues to use Li-heparin plasma for D-Dimer quantification, you are required to confirm all results between 0.5 g FEU/mL and 0.675 g FEU/mL with another D-Dimer method or with a citrated plasma sample. " If your laboratory continues to use Li-heparin plasma for D-Dimer quantification and chooses to confirm results with another method, you may request credit for that testin
Quantity5,263 kits
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062203TINA-QUANT D-DIMER TEST SYSTEM
Roche Diagnostics Corp., Mar 14, 2007