Recall Z-0400-2024 ยท FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days

Becton Dickinson Infusion Therapy Systems Inc.

BD Insyte Autoguard BC Shielded IV Catheter with Blood Control Technology 20GA X 1.00in, REF: 382533

DateNov 24, 2023 (initiated Oct 4, 2023)
StatusOpen, Classified
ReasonOver-the-needle, intravascular catheters may have needles that do not retract properly into the barrel, which could result in contaminated needlestick injury, blood leaking from the insertion site or catheter, and repeat catheter insertion attempts, due to catheter being inadvertently removed from the insertion site.
Root causeNonconforming Material/Component
ActionOn 10/4/23, recall notices were emailed to customers and distributors who were asked to do the following: 1) Destroy all unused product subject to the recall following your institution's process for destruction. 2) For indwelling catheters, no further action is needed. 3) Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 4) Complete and return the customer response form via email to BDRC48@bd.com Customers with recall questions, product complaints, or technical questions can call 1-844-823-5433, say recall
Quantity528,000
DistributionUS Nationwide distribution in the states of CA, NC, MI, WA, CO, FL, NJ, NM, IL, KY, TN, OH, TX, IN, LA, IA, MS, MO, NE, MD, NY, PA, GA, VA, ME, SC, RI, MA, PR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201075BD Insyte Autoguard Shielded IV Catheter, BD Insyte Autoguard BC Shielded IV Catheter, BD Insyte Autoguard BC Pro Shield
Becton, Dickinson and Company, Dec 18, 2020