ClearanceBrief
Recall Z-0350-2024 ยท FPA Set, Administration, Intravascular

Becton Dickinson & Company

BD gravity sets (Product Name, Catalog #) 32-IN GRAV SET W/0.2 MICRON FILTER BAG ACCESS PORT SS VALVE TEXIUMTM LOW ABSORPTION, Catalog # 4030B-07T and BD GRAVITY BLOOD SET 15 DROP 180 MICRON FILTER SmartSiteTM VALVE NONVENTED, Catalog # 42081E.

DateNov 17, 2023 (initiated Oct 11, 2023)
StatusOpen, Classified
ReasonAffected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).
Root causeError in labeling
ActionBD sent initial letters for this recall on October 11, 2023. A second notification was made available to customers on November 1, 2023 providing customers with additional information related to risk and recommended customer actions. In addition, the customer notification identified one additional product code that is in scope of this advisory. BD has identified that the infusion sets listed in Appendix 1 of the customer notification contain di(2-ethylhexyl) phthalate (DEHP) and have not been labelled accordingly. This could lead to the potential risk of DEHP exposure that may have short-term
DistributionWorldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WI, WY, Puerto Rico, Guam; and the cou
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022209ALARIS MEDICAL INFUSION SYSTEM ADMINISTRATION SETS
Alaris Medical Systems, Inc., Aug 6, 2002
K882302IVAC(R) UNVENTED VOL. INFUSION SET MODEL 52080
Y, Jul 12, 1988