ClearanceBrief
Recall Z-0233-2024 ยท FPA Set, Administration, Intravascular

Unomedical A/S

VariSoft infusion set, Model Number 1002828

DateNov 21, 2023 (initiated Oct 11, 2023)
StatusOpen, Classified
ReasonThe Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.hyperglycemia). Diabetic ketoacidosis.
Root causeProcess change control
ActionUnomedical a/s issued an Urgent Field Safety Notice to its sole consignee on 10/11/2023 via email. The notice explained the issue, potential risk to patient, and requested the distributor to quarantine all affected product pending return and to notify their customers.
Quantity1679 units
DistributionUS Nationwide distribution in the state of CA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162812Unomedical Comfort Subcutaneous Infusion Set, Unomedical Comfort Short Subcutaneous Infusion Set, Unomedical Neria Soft
Unomedical A/S, Nov 1, 2017