ClearanceBrief
Recall Z-0353-2024 ยท NSU Instrumentation For Clinical Multiplex Test Systems

BioFire Diagnostics, LLC

FilmArray TORCH REF HTFA-ASY-0104. HTFA-ASY-0104 is the BIOFIRE TORCH Base which includes barcode scanner, touch screen computer and USB ports. The BioFire FilmArray Torch (BioFire Torch) is an automated in vitro diagnostic (IVD) device intended for use with FDA-cleared or approved IVD BioFire FilmArray Panels. Additional Part/Reference Numbers as part of the system: - HTFA-ASY-0102: BIOFIRE TO

DateNov 20, 2023 (initiated Oct 17, 2023)
StatusOpen, Classified
ReasonDue to an increased risk of degradation of the power entry switch resulting from arcing inside of the power switch that may result in carbon build-up on the switch contacts. The carbon build-up may lead to excess heating inside of the power switch, subsequently leading to deformation of the power switch case. The deformation can create an opportunity for a power switch electrical short. In addition to an electrical short, deformation may result in an open circuit causing the power switch to fail. This event would only occur after the product is in use.
Root causeDevice Design
ActionOn 10/17/2023, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via FedEx overnight informing customers that bioMerieux has identified an increased risk of degradation of the poser entry switch of the TORCH base. This increase risk may be seen on systems that are power cycled (shutting down the system and then turning it on again) frequently. When turning on BIOFIRE TORCH systems, arcing inside of the power switch may result in carbon build-up on the switch contacts. The carbon build-up may lead to excess heating inside of the power switch, subsequently leading to deformation of the
Quantity5,406 systems
DistributionWorldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, N, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, Virgin Island, VT, WA, WI, WV, and W
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.