ClearanceBrief
Recall Z-0447-2024 ยท IYE Accelerator, Linear, Medical

Elekta, Inc.

ELEKTA AXESSE, REF XRT 2181, XRT 27891; Medical Charged-Particle Radiation Therapy System Accelerator, Linear

DateDec 1, 2023 (initiated Oct 25, 2023)
StatusOpen, Classified
ReasonElekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
Root causeUnder Investigation by firm
ActionElekta released Field Safety Notice to its consignees on the 10/25/2023 via email. The notice explained the issue, potential risk, and provides information for the safe use of the instrument until Elekta Field Service Representatives can perform the correction.
Quantity47 units
Distributionworldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K210500Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HD
Elekta Limited, Jun 11, 2021