Recall Z-0349-2024 ยท FPA Set, Administration, Intravascular
Becton Dickinson & Company
BD Alaris Pump infusion sets (Product Name, Catalog #) BD AlarisTM Pump Infusion Set Vented syringe adapter Smallbore Tubing, Catalog # 10010483; BD AlarisTM Pump Infusion Set 1.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-site, Catalog # 10010453; BD AlarisTM Pump Infusion Set 0.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-site, Catalog # 10010454; BD AlarisTM
| Date | Nov 17, 2023 (initiated Oct 11, 2023) |
| Status | Open, Classified |
| Reason | Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP). |
| Root cause | Error in labeling |
| Action | BD sent initial letters for this recall on October 11, 2023. A second notification was made available to customers on November 1, 2023 providing customers with additional information related to risk and recommended customer actions. In addition, the customer notification identified one additional product code that is in scope of this advisory. BD has identified that the infusion sets listed in Appendix 1 of the customer notification contain di(2-ethylhexyl) phthalate (DEHP) and have not been labelled accordingly. This could lead to the potential risk of DEHP exposure that may have short-term |
| Distribution | Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WI, WY, Puerto Rico, Guam; and the cou |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K022209 | ALARIS MEDICAL INFUSION SYSTEM ADMINISTRATION SETS Alaris Medical Systems, Inc., Aug 6, 2002 |