Recall Z-0446-2024 ยท IYE Accelerator, Linear, Medical

Elekta, Inc.

ELEKTA LIMITED Elekta Infinity, REF XRT 0511, XRT 0521, XRT 0531, XRT 0541; Medical Charged-Particle Radiation Therapy System Accelerator, Linear

DateDec 1, 2023 (initiated Oct 25, 2023)
StatusOpen, Classified
ReasonElekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
Root causeUnder Investigation by firm
ActionElekta released Field Safety Notice to its consignees on the 10/25/2023 via email. The notice explained the issue, potential risk, and provides information for the safe use of the instrument until Elekta Field Service Representatives can perform the correction.
Quantity291 units
Distributionworldwide
FDA recordOpen on fda.gov