Recall Z-0237-2024 ยท LLZ System, Image Processing, Radiological

CHANGE HEALTHCARE CANADA COMPANY

Change Healthcare Radiology Solutions, software versions 14.0, 14.1, and 14.2 Medical image and information management software.

DateNov 6, 2023 (initiated Oct 12, 2023)
StatusOpen, Classified
ReasonChange Healthcare has identified a software issue where Change Healthcare Radiology Solutions 14.0, 14.1, and 14.2 may not update records in external applications.
Root causeSoftware design
ActionField Safety Notice was distributed to all consignees on10/12/2023 by email. The notice explained the issue, potential risk to the patient, and directed users to to check the CHRS Exception Tracker on a regular basis, as failed updates will be listed with details on the failure. Recommended actions to take on potentially affected data: Impacted data can be identified by using the CHRS Exception Tracker. Customers are advised to review records of recent updates from CHRS to external systems and manually update any changes to study records that may have been missed. Change Healthcare will i
Quantity282 units
DistributionUS, Canada, Ireland, Saudi Arabia, France, UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140909MCKESSON RADIOLOGY
Mckesson Medical Imaging Company, Jul 2, 2014