ClearanceBrief
Recall Z-0369-2024 · BTM Ventilator, Emergency, Manual (Resuscitator)

SunMed Holdings, LLC

Medline manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Adult Resuscitator REF #s: CPRM1116FPM, CPRM1116FPM4, CPRM1116M, CPRM1116MC, CPRM1116PM, CPRM1116PMT, CPRM1126FM, CPRM1126FPM, CPRM1126FPWM, CPRM1126M; 2. Infant Resuscitator REF #s: CPRM3312FM, CPRM3312FM123, CPRM3312FPM, CPRM3312FPM3, CPRM3312FPMC, CPRM3312M, CPRM3312M3, CPRM3312P3M3, CPR

DateNov 21, 2023 (initiated Oct 11, 2023)
StatusOpen, Classified
ReasonA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
Root causeDevice Design
ActionCustomer notification was emailed October 11, 2023 advising customers that monitoring of tidal carbon dioxide in critically ill patients is recommended because use of accessories (e.g. PEEP valve) on the exhalation port of manual resuscitators may increase backward leak in the integrated manometer due to increased expiratory resistance. This affects all lots with duckbill-style patient valve with integrated manometer manufactured since 09/21/2019. Customers are to disseminate this information to all users and complete and return the response for. The notice also notes manual resuscitator IFUs
DistributionUS Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012842V*CARE MANUAL RESUSCITATOR
Ventlab Corp., Sep 18, 2001