Recall Z-0423-2024 ยท JGS Electrode, Ion Specific, Sodium

Siemens Healthcare Diagnostics Inc

RAPIDPOINT 500 Blood Gas System

DateDec 1, 2023 (initiated Sep 26, 2023)
StatusOpen, Classified
ReasonThere are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The erroneous sodium levels with positive bias may lead to unrecognized hyponatremia and/or the inappropriate treatment of hypernatremia may lead to iatrogenic hyponatremia, resulted in a delay in patient diagnosis and optimal patient management.
Root causeUnder Investigation by firm
ActionThe firm issued an Urgent Medical Device Correction notice on 9/26/2023 to customers asking that customers review the notice with the facility's Medical Director to determine the appropriate course of action for previously generated results, complete and return the Field Correction Effectiveness Check Form within 30 days, forward this notice to customers if your firm is a distributor, and retain the notification in your system's Operator's Guide and forward to those who may have received product. The Field Correction Effectiveness Check Form is to be returned via email to uscctsfcaecfax.team@s
Quantity12,363 units
DistributionDomestic: Nationwide Distribution. Foreign: Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Con
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113216RAPIDPOINT 500 BLOOD GAS ANALYER
Siemens Healthcare Diagnostics, May 3, 2012