Recall Z-0567-2024 ยท MKJ Automated External Defibrillators (Non-Wearable)

Physio-Control, Inc.

AC Power Adapter for use with the LIFEPAK 15 Monitor/Defibrillator, REF: 11140-000098

DateDec 15, 2023 (initiated Nov 6, 2023)
StatusOpen, Classified
ReasonAC Power Adapter (ACPA) may not charge the monitor/defibrillator batteries and may not allow the monitor/defibrillator to power on while plugged into the ACPA.
Root causeComponent change control
ActionOn 11/06/23, Recall and Field Safety Alert notices were sent to customers. The recall notices asked customers to do the following: 1) Complete and return business reply form by email to RSRecall@stryker.com 2) Firm recommends continued use. Follow the OI recommendations for daily inspection and testing according to the checklist provided in the Operating Instructions. a. The checklist includes disconnecting the defibrillator from the power adapter, turning the device on, and confirming that two batteries are installed and charged. Then, the defibrillator should be reconnected to the power a