Recall Z-0356-2024 ยท NZU Collector, Urine, Powered, Non Indwelling Catheter

Sage Products Inc

stryker Sage PrimaFit External Urine Management System for the Female Anatomy

DateNov 21, 2023 (initiated Oct 23, 2023)
StatusOpen, Classified
ReasonA small percentage of devices in three lots of product may contain a tape with a natural rubber latex adhesive which is not declared on the label
Root causeProcess change control
ActionStryker issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 10/23/2023 via USPS Certified Mail. The notice explained the issue and the potential risk to user and requested the following actions: 1. Locate the product listed above and if tape is detected on the product once the product is opened, do not use on patients that have a latex allergy. Isolate and discard any impacted product with tape to avoid accidental use by a patient with a latex sensitivity. 2. Distribute this notice to all hospital staff that need to be aware of the problem with this device. 3. Maintain a
Quantity174,960 devices
DistributionUS and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.