Recall Z-0334-2024 ยท FDF Colonoscope And Accessories, Flexible/Rigid

Aizu Olympus Co., Ltd.

Gastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, GIF-2TH180, GIF-H170, GIF-H180, GIF-H180J, GIF-H190, GIF-HQ190, GIF-LV1, GIF-N180, GIF-Q150, GIF-Q160Z, GIF-Q180, GIF-XP180N, GIF-XP190N, GIF-XTQ160. 510(k) Numbers: K112680, K954451, K051645, K100584, K111756, K011151,

DateNov 17, 2023 (initiated Sep 25, 2023)
StatusOpen, Classified
ReasonChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Root causeUnder Investigation by firm
ActionOlympus America issued Urgent Medical Device Correction letter on Sept. 25, 2023 to Endoscopy, Respiratory, Pulmonary and Urology Departments, Risk Management Department. Letter states reason for recall, health risk and action to take: Our records indicate that Olympus has serviced your flexible endoscope at one of our Olympus repair facilities. Following repair activities, Olympus uses a water flush of endoscope channels as part of our final endoscope channel inspection, and we subsequently dry the endoscope channels with compressed filtered air. Olympus discovered that our channel air drying
Quantity3627 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011151EVIS EXERA GASTROINTESTINAL VIDEOSCOPE GIF-Q160Z
The Olympus Optical Co., May 15, 2001
K051645EVIS EXERA 160A SYSTEM
Olympus Medical Systems Corporation, Oct 13, 2005
K100584EVIS EXERAII 180 SYSTEM
Olympus Medical Systems Corporation, Jul 2, 2010
K111756V1 SYSTEM
Olympus Medical Systems Corp., Mar 20, 2012
K112680EVIS EXERA III VIDEO SYSTEM
Olympus Medical Systems Corporation, Feb 16, 2012
K954451EVIS 140 SYSTEM
Olympus America, Inc., Mar 29, 1996