ClearanceBrief
Recall Z-0329-2024 ยท FET Endoscopic Video Imaging System/Component, Gastroenterology-Urology

Aizu Olympus Co., Ltd.

Colonoscope: for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve). Model Numbers: CF-2T160I, CF-2T160L, CF-H170I, CF-H170L, CF-H180AI, CF-H180AL, CF-H180DI, CF-H180DL, CF-H190I, CF-H190L, CF-HQ190I, CF-HQ190L, CF-LV1I, CF-LV1L, CF-Q150I, CF-Q150L, CF-Q160S, CF-Q180AI, CF-Q180AL. 510(k) Numbers: K954451, K1126

DateNov 17, 2023 (initiated Sep 25, 2023)
StatusOpen, Classified
ReasonChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Root causeUnder Investigation by firm
ActionOlympus America issued Urgent Medical Device Correction letter on Sept. 25, 2023 to Endoscopy, Respiratory, Pulmonary and Urology Departments, Risk Management Department. Letter states reason for recall, health risk and action to take: Our records indicate that Olympus has serviced your flexible endoscope at one of our Olympus repair facilities. Following repair activities, Olympus uses a water flush of endoscope channels as part of our final endoscope channel inspection, and we subsequently dry the endoscope channels with compressed filtered air. Olympus discovered that our channel air drying
Quantity3344 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001241EVIS EXERA COLONOVIDEOSCOPE CF-Q160 AL/I AND PCF-160 AL/I
The Olympus Optical Co., May 9, 2000
K051645EVIS EXERA 160A SYSTEM
Olympus Medical Systems Corporation, Oct 13, 2005
K100584EVIS EXERAII 180 SYSTEM
Olympus Medical Systems Corporation, Jul 2, 2010
K111756V1 SYSTEM
Olympus Medical Systems Corp., Mar 20, 2012
K112680EVIS EXERA III VIDEO SYSTEM
Olympus Medical Systems Corporation, Feb 16, 2012
K954451EVIS 140 SYSTEM
Olympus America, Inc., Mar 29, 1996