Recall Z-0284-2024 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Philips Ultrasound, Inc.

Philips Affiniti 70, Diagnostic Ultrasound System Model # 795210

DateNov 9, 2023 (initiated Oct 9, 2023)
StatusCompleted
ReasonTheir is the potential that diagnostic ultrasound systems operating with software version 10.0 may present a delayed image instead of a real time image.
Root causeSoftware design
ActionOn October 9, 2023, Philips issued a Urgent Medical Device Correction notification to affected consignees. Philips asked consignees to take the following actions: 1. Identify if your firm has an affected product. 2. In the 2D tab in the system touchscreen, disable the AutoSCAN feature every time a transducer or a Transducer preset is selected to prevent the system from potentially experiencing image lag issue. The AutoSCAN feature is disabled when the button is not highlighted in amber. 3. Please complete and return the attached response form to Philips promptly and no later than 15 day
Quantity7 systems
DistributionUS: AZ, CA, CT, FL, GA, ID, IL, KY, MD, MT, NE, NV, OH, TN, TX, UT, VA OUS: Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K231190EPIQ Series Diagnostic Ultrasound System
Philips Ultrasound, LLC, May 12, 2023