ClearanceBrief
Recall Z-0245-2024 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Stryker Corporation

Disposable Spetzler-Malis Bipolar forceps, 18CM IM DISP BAYONET 0.5MM TIP, Catalog Number 6770-180-005

DateNov 7, 2023 (initiated Sep 11, 2023)
StatusOpen, Classified
ReasonAn error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
Root causeUnder Investigation by firm
ActionAn URGENT Medical Device Recall notification letter was sent to customers on 9/11/23. Actions to be taken 1. Immediately check your inventory to locate and quarantine any affected products found at your facility. 2. Sign and return the enclosed Business Response Form (BRF) by email to Instruments.recalls@stryker.com, even if affected product is no longer in inventory. 3. Upon receipt of the completed BRF, a shipping label will be provided to return any recalled product on-hand and Stryker will contact you to arrange for replacement of your product(s). 4. Maintain awareness of this communicati
Quantity2,321 total units
DistributionUS Nationwide. Australia, Canada, Japan, Korea, Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121426AYNERGETICS DISPOSABLE SPETZLER-MALIS STANDARD BIPOLAR FORCEPS
Synergetics, Inc., Jul 10, 2012