Recall Z-0288-2024 ยท FDF Colonoscope And Accessories, Flexible/Rigid

Olympus Corporation of the Americas

EVIS EXERA III Gastrointestinal Videoscope

DateNov 9, 2023 (initiated Oct 2, 2023)
StatusOpen, Classified
ReasonSterilization failures when devices are sterilized per the Reprocessing Manual due to lack of drying time for the endoscope channel.
Root causeReprocessing Controls
ActionThe firm distributed recall notifications via mail to consignees dated 10/02/2023. Consignees are asked to inspect inventory for affected devices, ensure that all reprocessing personnel are aware and knowledgeable of the recall notice and its contents, and acknowledge receipt of the recall notification by returning the enclosed response form via email to olympus5381@sedgwick.com or by fax at 800-913-4345. Customers with any questions about the acknowledgement form are to call 866-912-9544. Customers with any questions about the recall are to contact Cynthia Ow by phone at 647-999-3203 or by em
Quantity1687 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112680EVIS EXERA III VIDEO SYSTEM
Olympus Medical Systems Corporation, Feb 16, 2012