Recall Z-0706-2024 ยท DPS Electrocardiograph

Baxter Healthcare Corporation

XScribe Cardiac Stress Testing System, version 6, with below product names and codes: a) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-BXXXX; b) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-DXXXX; c) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-XXXXX; d) XS6 WAM TTL KIT IEC XML BRAZIL, XSCRIBE-5AX-XXBRZ; e) XS6 WAM TTL SYS Z200 AHA XML, XSCRIBE-6AA-AAAAA; f) XS6 WAM TTL SYS Z200 BCRT IEC, XSCRIBE-6AA-ADBXX; g) XS6 WAM TTL SYS Z

DateJan 10, 2024 (initiated Dec 8, 2023)
StatusOpen, Classified
ReasonPotential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.
Root causeSoftware design
ActionBaxter issued an Urgent Medical Device Correction notice to its consignees on 12/08/2023 via USPS first class mail. The notice explained the problem, hazard involved, requested the following actions be taken until a software update becomes available: - Clinicians should evaluate the potential impact of the SCF as described to determine if it should be enabled or disabled during stress testing and provided the method for disabling the setting. - If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them, infor
Quantity2985 units
DistributionUnited States (including Puerto Rico), Canada, Australia, Bangladesh, Brunei, Chia, Hong Kong, India, Indonesia, Korea, Malaysia, New Zealand, Philippines, Singapore, Sri Lanka, Thailand, Algeria, Angola, Austria, Bahrain, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152944Xscribe Stress Exercise Testing Systems, Q-Stress Stress Exercise Testing Systems
Mortara Instrument, Inc., Jan 6, 2016