Recall Z-0705-2024 ยท DPS Electrocardiograph

Baxter Healthcare Corporation

Q-Stress Cardiac Stress Testing System, version 6, with below product name / product codes: a) QS6 AM12Q KIT AHA XML / QS6-BXXCS: b) QS6 AM12Q KIT BCRT IEC / QS-6CC-DGCXX: c) QS6 AM12Q KIT Z200 BCRT IEC / QS-6CC-DFCXX: d) QS6 AM12Q RS SYS AHA XML / QS-6JB-AXAXA: e) QS6 AM12Q SYS Z200 PCRT AHA XML / QS6-BLXCX: f) QS6 AM12Q TMX SYS Z200 PCRT AHA XML / QS6-BLXC1: g) QS6 AM12Q TTL KIT AHA / QS-

DateJan 10, 2024 (initiated Dec 8, 2023)
StatusOpen, Classified
ReasonPotential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.
Root causeSoftware design
ActionBaxter issued an Urgent Medical Device Correction notice to its consignees on 12/08/2023 via USPS first class mail. The notice explained the problem, hazard involved, requested the following actions be taken until a software update becomes available: - Clinicians should evaluate the potential impact of the SCF as described to determine if it should be enabled or disabled during stress testing and provided the method for disabling the setting. - If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them, infor
Quantity4687 units
DistributionUnited States (including Puerto Rico), Canada, Australia, Bangladesh, Brunei, Chia, Hong Kong, India, Indonesia, Korea, Malaysia, New Zealand, Philippines, Singapore, Sri Lanka, Thailand, Algeria, Angola, Austria, Bahrain, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152944Xscribe Stress Exercise Testing Systems, Q-Stress Stress Exercise Testing Systems
Mortara Instrument, Inc., Jan 6, 2016