ClearanceBrief
Recall Z-0095-2024 ยท GEX Powered Laser Surgical Instrument

Advance-Esthetic LLC

MED-810A Zemits NDPrime Laser products

DateOct 13, 2023 (initiated Oct 9, 2023)
StatusOpen, Classified
ReasonAdvance-Esthetic LLC failed to comply with the applicable standards regarding Electronic Product Radiation Control (EPRC) regulations associated with the laser product performance standards and certification. This Recall provides correction by the firm regarding Zemits NdPrime Yag Tattoo Removal Laser. The product requires update or is missing the following in part: remote interlock connector, safety interlocks, manual reset mechanism, beam attenuator, tests for determination of compliance, certification, class IV designation and warning, labels, and labeling.
Root causeRadiation Control for Health and Safety Act
ActionNotification letters will be sent to affected customers identifying the reason for notification and action to be taken. The manufacturer will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. For any queries or require further assistance, our customer support team is available at 888-999-3996 or customer-support@advance-esthetic.us.
Quantity19
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.