Recall Z-0221-2024 ยท IQA System, Environmental Control, Powered

Securitas Healthcare LLC

Arial Water Resistant Pendant with Wristband, Models: 59361 and 2560-59361; mobile patient alarm

DateNov 3, 2023 (initiated Sep 25, 2023)
StatusOpen, Classified
ReasonPremature battery depletion presents a risk that alarms will not be received by the Arial Wireless Emergency Call System.
Root causeComponent change control
ActionSecuritas Healthcare issued a recall notice to its consignees on 09/25/2023 via USPS, registered mail. The notice explained the issue, potential hazard, and requested the following actions: 1. Immediately replace Pendant battery upon Low Battery Alert 2. Customer should check Pendant battery status in Arial System application at least daily 3. Customer should check for Pendants that are no longer communicating with the Arial System at least daily This notice must be passed on to those who need to be aware within your organization. - Securitas Healthcare will be replacing all Pendants from
Quantity2691 units
DistributionWorldwide distribution - US Nationwide and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.