Recall Z-0157-2024 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Smith & Nephew, Inc.

JOURNEY II BCS ARTICULAR INSERT, LEFT 10 MM SIZE 5-6, REF 74027262; knee prosthesis

DateOct 24, 2023 (initiated Sep 27, 2023)
StatusOpen, Classified
ReasonThe JRNY II BCS XLPE ART ISRT SZ 5-6 LT 10MM was laser etched, labeled and packaged as a JRNY II BCS CNSTRD ART ISRT 5-6 LT 10MM. And also, a JRNY II BCS CNSTRD ART ISRT 5-6 LT 10MM was laser etched, labeled and packaged as a JRNY II BCS XLPE ART ISRT SZ 5-6 LT 10MM.
Root causeComponent change control
ActionSmith & Nephew, Inc. issued an Urgent Medical Device Recall Notice to its consignees on 09/27/2023 via email and overnight mail. The notice explain the issue, potential risk, and requested the following actions: Inspect inventory, locate, and quarantine the affected devices. Sales representatives, district offices, or distributors were directed to notify their customers The firm is seeking return of unused products.
Quantity28 units
DistributionWorldwide distribution - US Nationwide and the country of Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111711JOURNEY II BCS KNEE SYSTEM
Smith & Nephew, Inc., Sep 16, 2011
K140555JOURNEY II BCS CONSTRAINED INSERTS
Smith & Nephew, Inc., May 29, 2014