Recall Z-0209-2024 ยท GXD Generator, Lesion, Radiofrequency

Abbott

Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning generator

DateNov 2, 2023 (initiated Sep 28, 2023)
StatusOpen, Classified
ReasonCertain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.
Root causeProcess change control
ActionAbbott issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 09/28/2023 via FedEx 2nd day delivery. The notice explained the problem, risk, and requested return of the device for inspection. Those devices which pass, will be returned. Should your device(s) fail the inspection, your local Abbott Representative can advise the consignee of alternate commercially available products.
Quantity34 devices
DistributionUS, Taiwan, and Colombia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111576NT 2000 LESIONING GENERATOR
Neuro Therm, Inc., Sep 20, 2011