Recall Z-0183-2024 ยท FSO Microscope, Surgical, General & Plastic Surgery

Carl Zeiss Suzhou Co., Ltd.

OPMI LUMERA 300, REF 6137

DateOct 26, 2023 (initiated Aug 18, 2023)
StatusOpen, Classified
ReasonSurgical microscope for use in ophthalmology may be missing a screw from the suspension arm, which may cause the suspended components to fall from the suspension arm, which may injure any person under the suspended components.
Root causeProcess control
ActionOn 8/18/23, correction notices were mailed to customers who were asked to do the following: Before each use of the affected device, perform the following mandatory inspection: Refer to pictures in the notice showing a gap between the OPMI and the inspection arm. This gap should not be greater than 5mm. If the gap is greater than 5mm, please Immediately stop using the device! In this case the affected device can no longer be used. The firm is planning to inspect (repair if needed) the suspension arm immediately. Service staff will contact you to arrange an appointment for inspection (repa
Quantity8
DistributionUS Nationwide distribution in the states of MD, TX, CA, GA, LA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.