Recall Z-0213-2024 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Corin Ltd

Unity CR Femur Right, Size 6, Model Number 112.001.32 Primary and revision total knee arthroplasty

DateNov 2, 2023 (initiated Oct 13, 2023)
StatusOpen, Classified
ReasonPotential for Unity CR Inserts Right size 6 from batch 529803 to be incorrectly labelled as Unity CR Inserts Right size 7 from batch 532405 and visa versa.
Root causeProcess change control
ActionCorin issued a FIELD SAFETY NOTICE/RECALL LETTER to its consignees on 10/13/2023 via email. The notice explained the problem, potential risk if used, and requested the return of unused units. Distributors were directed to communicate the notice to applicable hospitals.
Quantity3 units
DistributionUS: MI, TX, IL