Recall Z-0170-2024 ยท NAY System, Surgical, Computer Controlled Instrument

Asensus Surgical Inc

Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Number X9007707), Software (SW) version 2.7.4; robotic surgical system

DateOct 30, 2023 (initiated Sep 15, 2023)
StatusTerminated
ReasonAsensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.
Root causeSoftware design
ActionAsensus Surgical issued an URGENT MEDICAL DEVICE RECALL to its consignees on 09/15/2023 by mail. The notice explained the problem, risk to health, and requested users to cease use of the device until software version 2.7.5 is installed.
Quantity21 units
DistributionUS: NJ, LA, FL, MN, IL, and Germany, Japan, Lithuania
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K223095Senhance Surgical System
Asensus Surgical, Inc., Mar 16, 2023