Recall Z-0195-2024 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Siemens Cios Alpha (VA30) mobile fluoroscopic C-arm X-ray System, designed for the surgical environment. 11105200

DateOct 27, 2023 (initiated Sep 12, 2023)
StatusOpen, Classified
ReasonSoftware issue for Cios Alpha, Cios Flow, and Cios Spin VA30 systems, fulfillment of new requirements from DIN and IEC standards for Cios VA30 systems required may result in patient misdiagnosis/repeat exposure
Root causeSoftware design
ActionSiemens Medical Solutions USA issued Urgent Medical Device Correction Letter (AX035/23/S).Letter states reason for recall, health risk and action to take. Siemens will correct the software error via Update Instruction AX036/23/S. With software version VA30K, the user interface of Cios VA30 systems will be changed to consider the current standards DIN 6862-3 and IEC 60601-2-54. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review,
Quantity237 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K210055Cios Alpha
Siemens Medical Solutions USA, Inc., Feb 5, 2021