ClearanceBrief
Recall Z-0202-2024 ยท CGA Glucose Oxidase, Glucose

Randox Laboratories Ltd.

RX Monaco-IVD Clinical Chemistry Analyser Catalog Number: RX5000

DateOct 31, 2023 (initiated Sep 6, 2023)
StatusOpen, Classified
ReasonRX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
Root causeUnder Investigation by firm
ActionRandox issued Medical Device Correction letter (REC687) on 9/13/23 to to US and PR customers on 9/12/23. Letter states reason for recall, health risk and action to take: Randox recommendations: " Auto rerun settings are disabled, see Appendix A for guidance for each Rx instrument. " Results with flag(<) reported as the bottom of the technical range. Sample Volume settings Low should not be applied. " Results with flag (>) reported as the top of the technical range. Sample Volume settings high and dilutions should not be applied. A manual dilution should be performed, and results multiplied by
Quantity13 units
DistributionUS Nationwide distribution including Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.