Recall Z-0141-2024 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

Smiths Medical ASD Inc.

Portex Pro-Vent Arterial Blood Sampling Kit with Dry Lithium Heparin for Gases and Electrolytes, REF 4599P-1

DateOct 20, 2023 (initiated Sep 6, 2023)
StatusOpen, Classified
ReasonThe complaint involved the absence of a component called "ASM Filter-Pro ABG SYR CAP" from the packaging of Pro-Vent Arterial Blood Sampling (ABG) Kits. Specifically, this issue affected a total of 31,900 kits under the FGSKU# 4599P-1 and Lot# 4331283. Tubes without caps have the potential to generate delayed or erroneous test results and/or exposure to infectious blood.
Root causeProcess control
ActionSmiths Medical issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 09/06/2023 by mail. The notice explained the problem with the device, potential risk, and requested the following: 1. Discontinue use, distribution, and quarantine all affected product in inventory. 2. Notify all potential users within the consignee organization 3. Distributors were directed to notify their customers. Questions: Customer Service 1-(800)-258-5361
Quantity31900 devices
DistributionUS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K952516PRO-VENT, LINE DRAW, PULSATOR
Smiths Industries Medical Systems, Inc., Jul 14, 1995