ClearanceBrief
Recall Z-0182-2024 ยท NWU Endoscope Introducer Kit

American Contract Systems, Inc.

Cath Cardiac Pack, REF IHCC03V

DateOct 26, 2023 (initiated Sep 1, 2023)
StatusOpen, Classified
ReasonDuring an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
Root causeProcess control
ActionOwens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 09/01/2023 by email. The notice explained the issue, risk, and requested the user discontinue the use of the I.V. Cath and Adhesive Dermabond and affix a warning label to each kit and case of affected product in inventory. The firm sent an updated letter on 09/22/2023 via email which corrected the product distribution date.
Quantity1260 units
DistributionUS Nationwide distribution in the states of FL, IA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.