Recall Z-0127-2024 ยท PGW Ear, Nose, And Throat Stereotaxic Instrument

Acclarent, Inc.

TruDi Navigation System, Model: FG-2000-00, Catalog: ENS022B, when used with TruDi Curette, Models: TDC0005Z and TDC0005

DateOct 20, 2023 (initiated Sep 21, 2023)
StatusOpen, Classified
ReasonWhen using affected curette and software, there is a discrepancy between the actual curette tip location and the location displayed on navigation systems intended for use during surgical procedures of the Ear Nose and Throat (ENT) and ENT skull base surgery, which may cause delayed/prolonged surgery, cerebrospinal fluid leak, visual impairment, or skull base structural damage.
Root causeSoftware Design Change
ActionOn 9/21/23, correction notices were mailed to customers who were asked to do the following: Please cease using the affected curette device with the affected navigation system until the new version of navigation system software version 2.3.2 is installed. 2. Please do not return any of the curettes within your inventory. 3. Complete and return the business response form via email to OneMD-Field-Actions@its.jnj.com 4. Forward this notice to any personnel in your facility that needs to be informed. 5. If any of the affected products have been forwarded to another facility, contact that facility
Quantity141
DistributionUS: CA, IL, WI, MA, TX, CO, FL, ID, KS, OH, OR, NE, MO, AZ, HI, TN, NY, ME, SC, NC, CT, DC, MN, LA, OK, PA, AL, IA, MI, SD
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K192397TruDi Navigation System
Biosense Webster, Jan 30, 2020