ClearanceBrief
Recall Z-0137-2024 ยท MWS Stabilizer, Heart

Maquet Cardiovascular, LLC

Maquet Acrobat-i Vacuum Stabilizer System Intended for use during performance of cardiac surgical procedures through a sternotomy incision approach. Product Code: OM-10000Z

DateOct 22, 2023 (initiated Sep 13, 2023)
StatusOpen, Classified
ReasonHousing Mount Jaw of these devices breaking during normal use and may result in injury to the epicardial tissue or vasculature, or a coronary artery or previously placed bypass graft, or could result in procedural delay and/or conversion.
Root causeDevice Design
ActionMaquet /Getinge issued Urgent Device Medical Device Removal (FSCA 2242352-09/06/2023-004-R) via FedEx with delivery receipt on 9/13/23. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any of the Acrobat-i Stabilizer System with the product codes/lot numbers listed in this notice. If you have unused/unexpired affected product that you will be returning from your inventory, please contact Maquet/Getinge Customer Service at 888-880-2874 between the hours of 6 AM and 5 PM Pacific Standard Time to request a return a
Quantity2951 units
DistributionNationwide Foreign: AT BE DE ES FR GB IT JP KR LU NL PL PT TR
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.