ClearanceBrief
Recall Z-0129-2024 ยท FPA Set, Administration, Intravascular

B. Braun Medical, Inc.

SA3000 VENTED MICRO SPIKE ADAPTER, REF No. 418121, for intravenous use.

DateNov 3, 2023 (initiated Sep 26, 2023)
StatusOpen, Classified
ReasonDevices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose both patients and healthcare workers to hazardous medication. There is also the potential for life-threatening situations to occur requiring additional medical intervention.
Root causeProcess control
ActionOn September 26, 2023, the firm notified customers via Urgent Product Recall Notification letters. Customers were informed of the potential for leakage at the male Luer lock site. Customers were instructed to report total amount of impacted product on hand to the firm, after which B Braun Customer Support will contact the customer with instructions on how to return the product for credit or replacement depending on the customer's need. Questions about this recall may be reported to BBMI s Postmarket Surveillance Department by calling 1-833-425-1464.
Quantity17,100 units
DistributionWorldwide distribution - US Nationwide and the countries of Canada and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K780551I.V. ADMINISTRATION SET WITH MINI-SPIKE
National Patent Development Corp., Apr 19, 1978