Recall Z-0145-2024 ยท FMF Syringe, Piston

Cardinal Health 200, LLC

Monoject 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack

DateOct 26, 2023 (initiated Sep 20, 2023)
StatusOpen, Classified
ReasonDemonstrated recognition and compatibility issues with syringe infusion pumps.
Root causeVendor change control
ActionCardinal Health notified customers on about 09/20/2023 via "URGENT MEDICAL DEVICE PRODUCT CORRECTION" letter sent FedEx next day delivery. Customers were instructed to: 1. REVIEW your inventory for the affected product codes and lots. Location of product code and lot are shown in above listed table and below example labeling. 2. COMMUNICATE with all personnel that utilize the Cardinal Health Monoject" Luer-Lock Tip syringes (1, 6, 12, 20, 35 and 60 mL) that they should not be used with syringe infusion pumps. 3. POST a copy of this notification in your storeroom where the product is stored
Quantity3,113,846 units
DistributionUS Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.