Recall Z-0143-2024 ยท LLZ System, Image Processing, Radiological

GE Healthcare

Universal Viewer Workflow Manager, Image processing radiological system

DateOct 23, 2023 (initiated Sep 8, 2023)
StatusOpen, Classified
ReasonWhen using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a third-party reporting application is used to launch exams, it could display information on a different patient than WFM and the Viewer.
Root causeSoftware design
ActionGE Healthcare notified consignees on about 09/08/2023 via letter. Consignees were instructed that until the software update is available the device can still be utilized through an updated browser setting or through discontinuing the use of bi-directional reporting. A GE Healthcare representative will contact each consignee to arrange for the correction. Consignees were instructed to complete and return the provided acknowledgement response.
Quantity57 units
DistributionNationwide distribution to AR, AZ, CA, CO, DE, FL, GA, IL, IN, ME, MI, MN, MO, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX. International distribution to Austria, Canada, France, Germany, Hong Kong, Israel, Italy, Philippines, Singapore, Thailand, Turkey, Vietnam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211312Universal Viewer
GE Healthcare, May 21, 2021