Recall Z-0194-2024 ยท EOQ Bronchoscope (Flexible Or Rigid)

Olympus Corporation of the Americas

EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-P190

DateNov 9, 2023 (initiated Oct 12, 2023)
StatusOpen, Classified
ReasonThere is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
Root causeUnder Investigation by firm
ActionAn "URGENT MEDICAL DEVICE CORRECTIVE ACTION" notification letter dated 10/12/23 was sent to customers. Actions to be taken by the end user: Our records indicate that your facility has purchased one or more of the affected bronchoscopes. Olympus requests you to take the following actions: 1. Inspect your facility for the referenced devices and ensure all personnel are completely knowledgeable and thoroughly aware of the Warnings in affected bronchoscope s Operation Manual for use with High-frequency therapy equipment and that Olympus high-frequency compatible bronchoscopes are compatible only
Quantity3948 devices
DistributionUS Nationwide distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201758EVIS EXERA III Bronchovideoscope Olympus BF-XP190, EVIS EXERA III Bronchovideoscope Olympus BF-P190
Olympus Medical Systems Corp., Jan 21, 2021