Recall Z-0069-2024 ยท GFD Dermatome

Zimmer Surgical Inc

Zimmer Dermatome Blades- 10-pack boxes of blades. Intended for use with the dermatome handpiece which is a skin grafting instrument that is intended to provide variable graft thickness and width capabilities Part number: 00880000010

DateOct 6, 2023 (initiated Sep 5, 2023)
StatusOpen, Classified
ReasonSkin grafts thin and non-uniform when using the affected blades. The issue would be identified at the time of use and may result in the need for additional harvests to adequately cover the area
Root causeProcess control
ActionZimmer issued Urgent Medical Device Recall Letter (FA 2023-00208 ) to Distributors and Risk Managers and Recall FAQs to affected consignees on September 5, 2023. Letter states reason for recall, health risk and action to take: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, immediately locate and quarantine affected product in your inventory. a. Your Zimmer Biomet sales representative may remove and return the affected product from your facility on your behalf. b. Alternatively, you may directly return
Quantity11,985 boxes (119,850 blades)
DistributionUS Nationwide Foreign: AUSTRALIA BRAZIL CHINA HONG KONG INDIA JAPAN MALAYSIA MEXICO NETHERLANDS ON SEOUL SINGAPORE TAIWAN THAILAND
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.